All Categories
enEN
ECG

Home / Products / Operation Room & ICU / ECG

6 Channel ECG Machine
6 Channel ECG Machine

6 Channel ECG Machine

Product Brochure:DOWNLOAD

  • Introduction
Introduction
1、Colorful graphic display of 12-lead ECG waveform on the same screen
2, button and touch screen synchronous operation, easy and practical
3, seven-inch color large TFT LCD display more comprehensive, more powerful functions
4, ECG automatic diagnosis and analysis function and independent choice of whether to print the analysis report
5、With lead off and key sound prompt function, real-time ECG waveform freeze function
6、Full keyboard Chinese input
7, manual, automatic, rhythm, physical examination and other modes of operation
8、Support SD high-capacity memory card, real-time large-capacity storage of various waveforms and cases
9, built-in rechargeable lithium battery, AC and DC dual-use
10, high-performance processor and integrated thermal array printing system
11, ultra-portable weight of only 1.2 kg, easy to carry out medical consultations
Dimension255mm×205mm×65mm
Weightabout 1.2kg
Display800×480 dots 7 inch color TFT LCD
OperationPush button + touch screen
Environmenttransport and StorageWork
Temperature-20℃~55℃5℃~40℃
Relative humidity25%~95%25%~85%
Atmospheric pressure strength70kPa ~106kPa 70kPa ~106kPa
Power supplyAC powersupply Rated voltage = 90V-240V
Rated frequency=50Hz/60Hz
Rated power = 35VA
DC power supply (built-in rechargeable lithium battery)Rated capacity = 2200mAh
Rated voltage = 10.8V
Discharge termination voltage ≥9V
Charging mode: constant current/constant voltage
Charging current (standard) = 0.2C5A (440mA)
Charging voltage (standard) = (13±0.1V)
Continuous maximum charging current = 2200mA
Continuous maximum discharge current = 2200mA
Cycle life ≥ 300 times
Power consumption35VA (max)
Fuse specificationT200mA 250V Ø5×20
RecorderRecording methodThermal dot matrix recording
Recording paper specificationsReel thermal recording paper
Recording paper width112mm
Effective recording width105mm
Paper speed10 mm/s, 25mm/s, 50mm/s (±3%)
Recording accuracy±5% (X-axis), ±5% (Y-axis)
Heart rate calculationCalculation methodPeak detection
Heart rate range30BPM~300BPM
Calculation accuracy±1 BPM
ECG host input methodFloating ground, defibrillation protection, pacing pulse suppression
Leads Standard12 leads, automatic lead change
Acquisition mode12-channel synchronization
A/D conversion12-bit / 16-bit / 24-bit
Measurement range> ±5mV
Time constant≥ 3.2s
Frequency response0.05Hz ~ 150Hz
Sensitivity2.5, 5, 10, 20 (mm/mV)
Input impedance50M (10Hz)
Input loop current≤ 50nA
Input voltage range< ±5 mVpp
Calibration voltage1mV±3%
Noise level< 15Vp-p
Inter-channel interference≤ 0.5mm
Patient leakage current< 10A (220V-240V/50Hz)
Patient auxiliary leakage current< 0.1A (DC)
Dielectric strength4000V rms
FilterAC filtering On/Off
Baseline drift filtering
Myoelectric interference filtering 25Hz, 35Hz, OFF
Low-pass filtering 150Hz/100Hz/75Hz
Common Mode Rejection Ratio (CMRR)>100dB